Direct analysis of potentially genotoxic impurities by thermal desorption-field asymmetric waveform ion mobility spectrometry-mass spectrometry

Analytical Methods Pub Date: 2013-07-11 DOI: 10.1039/C3AY40676H

Abstract

Thermal desorption has been combined with field asymmetric waveform ion mobility spectrometry and mass spectrometry for the rapid, direct analysis of isobaric potentially genotoxic impurities (PGIs) in a surrogate active pharmaceutical ingredient. FAIMS-selected PGIs were detected with limits of quantification <0.2 ppm, below the threshold of toxicological concern, with %RSD <8.4%, at the 1 ppm level.

Graphical abstract: Direct analysis of potentially genotoxic impurities by thermal desorption-field asymmetric waveform ion mobility spectrometry-mass spectrometry